This information is valid for the following devices: Fridge-tag 2L/Fridge-tag Ultra Low
The setting to remove or not the ALARM symbol from the display is a factory setting.
The possible settings are :
Option 1: Alarm indication "All Alarms"
With this option the alarms will be visible on the display with an alarm symbol for 30 days.
By pressing the READ button, the warning symbol ! will be disabled for the corresponding alarms. The alarm symbol cannot be cancelled nor reset.
Note
• In this mode only one upper and one lower alarm will be triggered per day.
• The alarm symbol will be present on the display for 30 days.
• The warning symbol ! can be deactivated by confirming all existing alarms in the readout mode.
• The alarm buzzer stops when the alarm is confirmed within the set alarm limits. Otherwise the buzzer pauses for approx. 1 hour and starts again for up to 168 hours (7 days).
Option 2: Alarm indication “unconfirmed alarms”:
The alarms are shown with the alarm symbol until all alarms (in the 30-day history) have been confirmed as solved by pressing the READ button. Afterwards the display will show the OK symbol until a new alarm is triggered.
By pressing the READ button the warning symbol ! will be disabled for the corresponding alarms. The alarm symbol disappears and the OK symbol will be shown again.