This information is valid for the following devices: FT2L/FT ULT
The PDF displays the total time out of bounds(Accumulated), this time doesn’t have to be in one piece while the alarm will be only triggered if the temperature violation is continuous (Single Alarm)
Example setup: lower temperature limit <–0.5°C, duration >1 hour
In the above example multiple low temperature deviations with exposure times of less than 1 hour occurred. The cumulative daily time below the limits adds up to about 3.5 hours but no alarm will be triggered. The same behaviour also applies to the upper alarm.
The ALARM is set to be a Single Alarm.